Review

Radiation Dose Optimization in Indian Computed Tomography: Current Diagnostic Reference Levels and Future Directions

1★, 2, 3, 4, 5, 6★ Corresponding author

  1. 1Medical Radiology and Imaging Technology, Jaipur National University, Jagatpura, Jaipur, Rajasthan, INDIA.
  2. 2Medical Radiology and Imaging Technology, Paramedical College, F/O Medicine, Aligarh Muslim University (AMU), Aligarh, Uttar Pradesh, INDIA.
  3. 3Medical Imaging Technology, Paramedical Sciences, Jamia Hamdard, New Delhi, INDIA.
  4. 4Medical Radiology and Imaging Technology, Rayat Bahra University, Mohali, Punjab, INDIA.
  5. 5Radiology and Imaging, Emversity, Indiranagar, Bangalore, Karnataka, INDIA.
  6. 6Department of Radiology, Galgotias University, Greater Noida, Uttar Pradesh, INDIA.

CORRESPONDENCE

Shagufta Parveen

Assistant Professor, Jaipur National University, Jagatpura, Jaipur, Rajasthan, INDIA.

shaguftatap395@gmail.com

Received: 28-07-2026; Revised: 03-09-2026; Accepted: 19-10-2026.

Volume 15, Issue 3 · pp. 641–647 · 2026 · DOI: 10.5530/ajbls.20260019

View on AJBLS original site ↗

ABSTRACT

CT Scan is important for both diagnosing and monitoring medical conditions. However, the increased use of CT has raised concerns about patient radiation exposure to optimize the radiation dose while maintaining the best diagnostic image quality, Diagnostic Reference Levels (DRL) must be established to define any unusual increases in radiation dose. This study examines CT DRLs in India. Published articles were reviewed in this work in Indian context on the adult Diagnostic Reference Levels for specific anatomic regions to identify the cause and extent of deviations and to examine possible strategies to optimize Computed Tomography radiation dose. Variations in radiation dose within and among Computed Tomography centers are frequently reported, including scanning technology, study design, and the use of computed tomography parameters. There are only few studies published on DRLs in India. These types of studies are essential for developing national standards in different regions of the country and to provide a valuable and correct comparison with international DRLs.

KEYWORDS

0% READ

FULL TEXT

INTRODUCTION

Computed Tomography is also called computed axial tomography, CT scan, or CAT scan. It is an accepted imaging method to evaluate the whole body. The Images are acquired directly in the axial plane of varying tissue thickness. The advantage of these tomographic images compared to conventional X-rays is that they contained detail information about specific parts of the cross-section through computer imaging. Some pathology can be seen in the sagittal and coronal planes by reconstruction of the images by the computer. CT has undergone many changes, and now multi-detector CT scanners have been evolved which have better application in clinical field. The development of helical and multi detector row arrangements has improved CT applications (Prokop, 2003). CT scans use ionizing radiation that can travel through the body and produce images of structures present inside the body. CT is associated with high radiation doses. As a result, cautious application of this imaging modality necessitates a strict commitment to radiation protection principles: justification, optimization, and minimization, ensuring that patient risk is not greater than the gained benefit from the technique (Tsapaki et al., 2006; Vetter, 2008). The methods and wide range of exposure parameters used by modern CT scanners can greatly affect the radiation dosage that is administered to the patient. The development of Multi-Detector Computed Tomography (MDCT), which allows for more detailed anatomical information and diagnosis with greater accuracy, revolutionized diagnostic medicine. This is accomplished via multi-planar reformation, isotropic volume data collection, and images that are less than 1 mm thick. Many radiographic and fluoroscopic exams are now replaced by MDCT. Among all the diagnostic modalities, CT examinations provide the highest radiation dose. For example, the radiation dose of a routine volumetric single-section chest CT technique is more than a hundred times of the total radiation dose from frontal and lateral chest radiograph. Over the past three decades, there has been a noticeable increase in the number of CT exams performed (Varghese et al., 2018). High-quality images should be captured during CT examinations that focus on specific clinical goals without giving an unnecessary dose to the patient. Therefore, ICRP suggested the idea of DRL and the priority is to optimize the patient's radiation dose.

This study reviewed established DRLs for CT examinations used in India. Our objectives were to ( 1) Analyze CT DRL in India (2) Identify factors that influence DRL variability, and (3) Explore strategies that can optimize CT radiation dose in the Indian health care context.

CT DOSE DESCRIPTORS

Dose Length Product (DLP)

A CT tube emits milli grays per centimeter as a Dose Length Product (DLP). DLP considers the radiation source's z-axis length (the patient's long axis) (Bushberg and Boone, 2011; Durand and Mahesh, 2012). It uses one phantom slice to monitor dosage. DLP=(CTDIVOL) x (length of scan, cm) (Figure 1).

The unit to measure DLP is mGycm.

Computed Tomography Dose Index (CTDI)

CT imaging measures radiation dose using the Computed Tomography Dose Index (CTDI). CTDI is a standard metric of CT scanner radiation dose output that lets users compare CT scanners. We quantify CTDI in mGy. Initial measurements included CTDI across a 100-mm ionization chamber (CTDI 100) and weighted readings. Nowadays, helical scanners use CTDIvol instead of CTDI100 and CTDIw (McNitt-Gray, 2002) (Figure 2).

Diagnostic Reference Level (DRL)

Optimizing with DRLs is feasible. diagnostic Reference levels are investigation levels used to optimize patient protection during diagnostic and interventional procedures. In medical imaging using ionizing radiation, it indicates whether a procedure's radiation dosage is too high or low for everyday regular conditions (Vañó et al., 2017). In 1989, the UK proposed national reference dosages for some frequent radiographic examinations after a mid-1980s national survey found significant variability in hospital practices for similar procedures (Kerr, 1991). In 1990, the International Commission on Radiological Protection (ICRP) introduced diagnostic reference levels and developed them with publication 73 and 105 and ICRP supporting advice 2; Diagnostic reference level is "a form of investigation level, apply to an easily measured quantity, usually the absorbed dose in air, or tissue-equivalent material at the surface of a simple standard phantom or representative patient," according to the ICRP (Harding and Thomson, 1997). Diagnostic radiology national and regional reference values are usually the 75th percentile of sample average doses. Because facilities in different countries and regions may have different equipment and procedures, DRLs must be updated constantly with technology updates (Figure 3).

TECHNOLOGICAL ADVANCEMENTS IN CT DOSE OPTIMIZATION

The DRL serves as the standard for keeping radiation exposure in medical imaging as low as possible without compromising the ability to provide important diagnostic data (ALARA). In CT imaging, various optimization strategies and techniques have been developed and implemented to reduce patient radiation dose while maintaining diagnostic image quality. Here we present some common strategies and their approximate CT dose reduction.

Automatic Exposure Control (AEC)

The AEC modulates tube current based on patient anatomy and size to deliver radiation only where needed. Compared to fixed tube current scanning, this method reduces dosage by 25-50% (McCollough et al., 2006) (Figure 4).

Tube voltage selection (kVp optimization)

Lowering tube voltage (kVp) reduces patient dose and maintains image quality. A 10 kVp reduction reduces dosage by 20-30% (Kalra et al., 2004).

Iterative reconstruction Algorithms

Iterative reconstruction reduces noise and artifacts, allowing lower radiation doses. These algorithms minimize dosage by 30-60% over filtered back projection (Willemink et al., 2013).

Pitch Adjustment

If the value of the pitch (feed revolutions per table) increases, it is possible to reduce the overlap of the received data, resulting in a dose reduction of 10-40%. However, in some cases, high pitch values can degrade image quality (McCollough et al., 2006).

ARTIFICIAL INTELLIGENCE APPLICATIONS IN CT

Comparative Analysis: Indian DRLs and International Benchmarks (McCollough et al., 2025).

Table 3:

Examination

India (South India) CTDIvol

India (South India) DLP

International (IAEA/EC/UK/ USA) CTDIvol

Head

27-47

620-1041

47-60

Chest (Non-contrast)

10

445

9.5-15

Chest (Contrast)

-

-

10-15

Abdomen

12-16

482-550

10.9-14

Pelvic

-

-

400-650

Cervical Spine

-

-

17-28

ARTIFICIAL INTELLIGENCE APPLICATIONS IN

CT is fast changing because to AI's ability to balance radiation dose reduction and image quality. AI has the potential to improve CT imaging, according to meta-analysis of five clinical validation trials with 929 participants between 2022 and 2024 (Zhang et al., 2025). Recent studies show that AI-based interventions, particularly using deep learning algorithms like CNNs, enhance image quality across multiple anatomical regions with a mean difference of 0.70 (95% CI 0.43-0.96; p<0.001), without affecting diagnostic accuracy.

AI-enhanced low-dose CT (LDCT) uses improved noise reduction and artifact correction to produce images comparable to standard-dose CT scans. This is especially useful in radiation-dose-minimizing applications like lung cancer screening and liver lesion identification, where diagnostic fidelity is crucial Machine learning techniques offer patient-specific dosage prediction based on scanner characteristics and demographics, allowing individualized scan protocol changes to maximize dose and picture quality. AI-driven workflow optimization automates protocol selection, patient placement, and real-time dosage monitoring in CT operations beyond picture reconstruction. In resource-limited Indian healthcare institutes like KMC Manipal, AI-enabled workflows improve patient safety and departmental efficiency, but infrastructure and training issues remain (Maleki Varnosfaderani & Forouzanfar, 2024). . Low-dose image enhancement, tailored dosage prediction, and workflow automation are key AI uses in CT that increase patient safety and diagnostic outcomes. Future study should focus on standardization and clinical integration for varied healthcare environments like India.

Policy and Regulatory Framework

To use Computed Tomography (CT) safely and effectively while avoiding radiation exposure in healthcare, strong laws and regulations are needed. Nationwide Radiation Protection Policies The Atomic Energy Regulatory Board (AERB) regulates dose limits, equipment certification, and licensing in India. Recent AERB National Diagnostic Reference Levels (DRL) guidelines emphasize dose optimization adapted to regional needs and align with ICRP and WHO standards, supporting global radiation safety goals Quality Control and Accreditation QA programs are becoming important for radiation dose monitoring and imaging standards. Hospital certification authorities like the National Accreditation Board for Hospitals and Healthcare Providers (NABH) require structured dose audits and equipment checks for radiation safety. QA measures improve diagnosis accuracy and patient safety Capability-building training.

Radiologists, technologists, and medical physicists must be trained to defend against radiation. National initiatives by AERB and organizations like the Indian Radiological and Imaging Association encourage dose optimization, DRL application, and future technologies like AI in CT dose control through workshops, certification, and CME. This continual training helps clinicians adapt to India's different clinical situations (Das et al., 2025).

Ethical and Clinical Considerations

Computed tomography ethics require balancing radiation exposure and diagnostic quality. Clinicians must keep doses as low as reasonably feasible (ALARA) without compromising diagnostic accuracy to avoid radiation exposure and get clinically meaningful information. Informed consent and collaborative decision-making require patient information regarding radiation risks, procedure advantages, and alternatives.

Radiologists, medical physicists, technicians, and referring physicians must work together to improve scan protocols, monitor dosages, and ensure patient safety. This partnership promotes diagnostic imaging safety and ethics (Brady et al., 2024).

Figure 1: DLP (Dose length product).
Figure 2: CTDI (Computed tomography dose index).
Table 1: National Diagnostic Reference Levels (NDRL) for CT currently used in different states of India (CTDI=mGy, DLP=mGy/cm).

State

Descriptor

Head

Chest

Abdomen

South India

Computed Tomography Dose Index (CTDIv)

47

10

12

Dose length product (DLP)

1041

445

550

Pudhucherry, India

Computed Tomography Dose Index (CTDIv)

32

12

16

Dose Length Product (DLP)

925

456

482

Kerala, India

Computed Tomography Dose Index (CTDIv)

27

5

9

Dose Length Product (DLP)

620

164

319

FUTURE PERSPECTIVES

Photon-counting CT reduces patient radiation exposure by improving spatial resolution and dosage efficiency over conventional CT detectors. Quantum computing is being studied to improve complex picture reconstruction and optimization to speed up and improve dose reduction. Big data analytics use large patient datasets to dynamically alter CT procedures for real-time dose monitoring and individualized radiation management (Das et al., 2024). For safety and efficacy, clinical validation of these sophisticated technologies and India-specific models that represent regional imaging practices and patient demographics should be prioritized in future research.

DISCUSSION

Standardizing diagnostic reference levels for operations in different modalities is one technique to reduce radiation dose. ( Livingstone and Dinakaran, 2011) conducted the first Indian DRL study in 2011. Regional DRL is created in Tamil Nadu, South India. This study was part of the AERB-funded CT dose measuring effort on 127 Tamil Nadu CT scanners from 2006 to 2008. A 32-cm Polymethylmethacrylate (PMMA) body phantom measured CTDI, which was utilized to create DLP for different body areas. The regional DRLs for the thorax, abdomen, and pelvis were 557, 521, and 294 mGy/cm. All scanners were quality-checked before regional dose estimate. CT thorax, abdomen, and pelvis mean effective doses were 8.04, 6.69, and 4.79 mSv, respectively, based on DLP values.

We included three articles in this review (Das et al., 2024; (Livingstone and Dinakaran, 2011). CT National Reference Levels (NDRL) for Indian states are shown in Table 1. In different parts of India, DRLs vary greatly. Kerala has a CT head CTDI of 27mGy, while South India has 47mGy. Equipment, protocol, and local practices vary by location nationwide. This discrepancy emphasizes the necessity for a standard NDRL approach in India. South India's study (Saravanakumar et al., 2016; Saravanakumar et al., 2015). included Tamil Nadu, Puducherry, Kerala, and Karnataka (110 CT scanners). All CT scanners in this investigation were single-slice and helical multidetector row (1-256 slices). In Puducherry (Saravanakumar et al., 2017) six CT scanners were used, while in Kerala (Huda and Mettler, 2011),16 were used in eight main cities out of 13. QA and QC were done on all equipment in these studies. Some CT scanners have higher CTDI and DLP than others, indicating higher radiation output due to operating parameters like pitch, kV, slice collimation, scan mode, slice thickness, measurement field, tube rotation, and couch increment. Periodic QA is recommended. CT scanners should choose slice thickness and scan parameters to reduce patient radiation.

Table 2: CTDI, DRL in pediatric CT (Huda and Mettler, 2011).

Study region

Age group

<1

1-5

Head

20 CTDIv

38 CTDIv

352 DLP

505 DLP

Chest

7 CTDIv

8 CTDIv

120 DLP

132 DLP

Abdomen

12 CTDIv

14 CTDIv

252 DLP

270 DLP
Figure 3: Optimization of radiation protection in digital radiology techniques.
Figure 4: AEC sensor (Automatic exposure control).

Indian DRL establishment research is also published. One study described Tamil Nadu's regional pediatric CT dose (Huda and Mettler, 2011) in Table 2. That study found that the 75th percentile of DLP of infant head for 0-1 year olds is greater and lower than European PiDRL for 1-5 year olds, and for abdominal and chest operations is higher than European pediatric DRL for both age groups. This is largely owing to long scan times for both procedures and age groups. The age and weight-based protocol for pediatric CT examination for DRL comparison is lacking, according to this analysis.

Another one-month study at seven Kerala tertiary care institutions examined multidetector CT radiation doses by anatomical location. Single-phase head exams had a 75th percentile CTDIv of 68.1 mGy and DLP of 1120 mGy/cm. The 75th percentile CTDIv and DLP for single-phase abdomen examinations were 10.6 and 509.3; multiphase examinations were 14.6 and 2666.9; single-phase chest examinations were 23.4 and 916.7; multiphase chest were 19.9, 1737.6; and single-phase neck examinations were 24.9, 733.6 and 2076. Iterative reconstruction and automated tube current modulation are standard on all MDCT scanners. Despite greater image quality, MSCT's smaller pitch factor increases CTDIv (CTDIv=CTDIw / pitch) (Huda and Mettler, 2011). DRL should be calculated independently for single-and multi-section research. Delay phases should be incorporated in multi-phase investigations.

CONCLUSION

The literature concludes that over past few years, the establishment of Diagnostic Reference Level in India have not considerably increase. There are only few studies published on DRL which is only from south India. There should be awareness and knowledge of DRL in public and private hospitals. Radiation safety in diagnostic radiology is important, not only because of regulatory requirements but also because of personal and patient consideration. Professionals must agree on the collection of DRL data to establish DRL values and optimize the dose. Current CT scanning technology uses two measures to estimate patient radiation exposure: CTDIv and DLP. In the current review, significant dose variations were noted, and these variations were due to differences in scanning protocols and scanning-related parameters. Standardization of protocols and optimizing exposure parameters are necessary. DRL is a useful tool for continuous improvement of clinical practice. Radiologists and technologists need to be aware of local dose variations within their department as compared with national and international DRLs.

ABBREVIATIONS

CT
Computed Tomography
CAT
Computed Axial Tomography
MDCT
Multi-Detector Computed Tomography
DRL
Diagnostic Reference Level
DLP
Dose Length Product
CTDI
Computed Tomography Dose Index
AEC
Automatic Exposure Control
kVp
Kilovoltage Peak
ALARA
As Low As Reasonably Achievable
AERB
Atomic Energy Regulatory Board
NABH
National Accreditation Board for Hospitals and Healthcare Providers
ICRP
International Commission on Radiological Protection
WHO
World Health Organization
PMMA
Polymethylmethacrylate
MSCT
Multi-Slice Computed Tomography
LDCT
Low-Dose Computed Tomography
CNN
Convolutional Neural Network
QA/QC
Quality Assurance/ Quality Control
NDRL
National Diagnostic Reference Level

CONFLICT OF INTEREST

The authors declare that there is no conflict of interest.

REFERENCES

As published

Showing references and in-text citations exactly as published.

  1. Brady, Z. (2024). Multidisciplinary approach to radiation protection in diagnostic imaging. Journal of Medical Imaging and Radiation Sciences, 55(2), 157–164.GOOGLE SCHOLAR
  2. Bushberg, J. T, & Boone, J. M. (2011). The essential physics of medical imaging (3rd ed.). Lippincott Williams & Wilkins.GOOGLE SCHOLAR
  3. Das, I. J. (2024). Big data and real-time dose monitoring in medical imaging: Opportunities and challenges. Medical Physics, 51(12), 7250–7265.GOOGLE SCHOLAR
  4. Das, S. K. (2025). Capacity building and training for radiation safety: Challenges and opportunities in india. Journal of Medical Physics, 50(1), 12–18.GOOGLE SCHOLAR
  5. Durand, D. J, & Mahesh, M. (2012). Understanding CT dose display. Journal of the American College of Radiology, 9(9), 669–671.GOOGLE SCHOLAR
  6. Harding, K, & Thomson, W. H. (1997). Radiological protection and safety in medicine-ICRP 73. European Journal of Nuclear Medicine, 24(10), 1207–1209.GOOGLE SCHOLAR
  7. Huda, W, & Mettler, F. A. (2011). Volume CT dose index and dose-length product displayed during CT: What good are they? Radiology, 258(1), 236-242.GOOGLE SCHOLAR
  8. Kalra, M. K, Maher, M. M, Toth, T. L, Schmidt, B, Westerman, B. L, Morgan, H. T, & Saini, S. (2004). Techniques and applications of automatic tube current modulation for CT. Radiology, 233(3), 649–657.GOOGLE SCHOLAR
  9. Kerr, I. H. (1991). Patient dose reduction in diagnostic radiology. Clinical Radiology, 43(1), 2–3.GOOGLE SCHOLAR
  10. Livingstone, R. S, & Dinakaran, P. M. (2011). Radiation safety concerns and diagnostic reference levels for computed tomography scanners in tamil nadu. Journal of Medical Physics, 36(1), 40–45.GOOGLE SCHOLAR
  11. McCollough, C. H. (2025). Diagnostic reference levels and achievable doses in medical imaging: Global trends and practical implications. Radiology, 294(1), 12–20.GOOGLE SCHOLAR
  12. McCollough, C. H, Bruesewitz, M. R, Kofler, J. M, & Jr (2006). CT dose reduction and dose management tools: Overview of available options. Radiographics, 26(2), 503–512.GOOGLE SCHOLAR
  13. McNitt-Gray, M. F. (2002). AAPM/RSNA physics tutorial for residents: Topics in CT: Radiation dose in CT. Radiographics, 22(6), 1541–1553.GOOGLE SCHOLAR
  14. Prokop, M. (2003). Multislice CT: Technical principles and future trends. European Radiology, 13(suppl 5), 3–13.GOOGLE SCHOLAR
  15. Saravanakumar, A, Vaideki, K, Govindarajan, K. N, & Jayakumar, S. (2014). Establishment of diagnostic reference levels in computed tomography for select procedures in Puducherry, india. Journal of Medical Physics, 39(1), 50–55.GOOGLE SCHOLAR
  16. Saravanakumar, A, Vaideki, K, Govindarajan, K. N, Devanand, B, Jayakumar, S, & Sharma, S. D. (2016). Establishment of CT diagnostic reference levels in select procedures in south india. International Journal of Radiation Research, 14(4), 341.GOOGLE SCHOLAR
  17. Saravanakumar, A, Vaideki, K, Govindarajan, K. N, Jayakumar, S, & Devanand, B. (2015). Estimation of dose reference levels in computed tomography for select procedures in Kerala, india. Journal of Medical Physics, 40(2), 115–119.GOOGLE SCHOLAR
  18. Saravanakumar, A, Vaideki, K, Govindarajan, K. N, Jayakumar, S, & Devanand, B. (2017). Assessment of regional pediatric computed tomography dose indices in tamil nadu. Journal of Medical Physics, 42(1), 48–54.GOOGLE SCHOLAR
  19. Tsapaki, V, Aldrich, J. E, Sharma, R, Staniszewska, M. A, Krisanachinda, A, Rehani, M, & Prokop, M. (2006). Dose reduction in CT while maintaining diagnostic confidence: Diagnostic reference levels at routine head, chest, and abdominal CT-IAEA- Coordinated research project. Radiology, 240(3), 828–834.GOOGLE SCHOLAR
  20. Vañó, E, Miller, D. L, Martin, C. J, Rehani, M. M, Kang, K, Rosenstein, M, & Rogers, A. (2017). ICRP publication 135: Diagnostic reference levels in medical imaging. Annals of the ICRP, 46(1), 1–144.GOOGLE SCHOLAR
  21. Varghese, B, Kandanga, I, Puthussery, P, Vijayan, D, Babu, S. H, Aneesh, M. K, & Kumar, S. (2018). Radiation dose metrics in multidetector computed tomography examinations: A multicentre retrospective study from seven tertiary care hospitals in Kerala, south india. Indian Journal of Radiology and Imaging, 28(2), 250–257.GOOGLE SCHOLAR
  22. Varnosfaderani, M, S, & Forouzanfar, M. (2024). The role of AI in hospitals and clinics: Transforming healthcare in the 21st century. Bioengineering, 11(4), 337. https://doi.org/10.3390/bioengineering11040337CrossrefGOOGLE SCHOLAR
  23. Vassileva, J, & Rehani, M. (2015). Diagnostic reference levels. AJR: American Journal of Roentgenology, 204(1), W1-w3.GOOGLE SCHOLAR
  24. Vetter, R. J. (2008). ICRP publication 103: The recommendations of the international commission on radiological protection. Health Physics, 95(4), 445–446. https://doi.org/10.1097/01.HP.0000324200.73903.5BCrossrefGOOGLE SCHOLAR
  25. Willemink, M. J, Jong, D, P, A, Leiner, T, Heer, D, L, M, Nievelstein, R. A, Budde, R. P, & Schilham, A. M. (2013). Iterative reconstruction techniques for computed tomography. Part 1: Technical Principles. European Radiology, 23(6), 1623–1631.GOOGLE SCHOLAR
  26. Zhang, S, Zhu, Z, Yu, Z, Sun, H, Sun, Y, Huang, H, Xu, L, & Wan, J. (2025). Effectiveness of AI for enhancing computed tomography image quality and radiation protection in radiology: Systematic review and meta-analysis. Journal of Medical Internet Research, 27, e66622. https://doi.org/10.2196/66622CrossrefGOOGLE SCHOLAR

Cite this article

SELECT FORMAT

Parveen, S., Arfat, M., Taiba, T., Kiran, K., Chauhan, K., & Singh, N. (2026). Radiation Dose Optimization in Indian Computed Tomography: Current Diagnostic Reference Levels and Future Directions. Asian Journal of Biological and Life sciences, 15(3), 641–647. https://doi.org/10.5530/ajbls.20260019