IJPI · Lens · Development and Optimization of a Validated UV Spectrophotometric for Simultaneous Estimation of Thymoquinone and Temozolomide

PDF

Article · Volume 16, Issue 4 · pp. 1629–1634

Development and Optimization of a Validated UV Spectrophotometric for Simultaneous Estimation of Thymoquinone and Temozolomide

Krutika Lolsure1, Vijay Kumbar1, Priya Shetti1, Mitali Kubade1, Malleswara Peram2

  1. 1Dr. P K Basic Science Research Center, KLE Academy of Higher Education and Research, Neharu Nagar, Belagavi, Karnataka, INDIA.
  2. 2Department of Pharmaceutics, Chebrolu Hanumaiah Institute of Pharmaceutical Science, University in Chowdavaram, Guntur, Andhra Pradesh, INDIA.

Abstract

Introduction Glioblastoma multiforme, an extremely aggressive form of brain cancer, presents significant challenges in treatment, despite the application of surgical intervention, radiation therapy, and chemotherapy. Researchers are investigating alternative therapeutic options, including phytochemicals such as Thymoquinone, which has demonstrated potential in targeting specific cancer pathways. The establishment of a dependable analytical method for the simultaneous quantification of Thymoquinone and Temozolomide is essential for enhancing research efforts and ensuring quality control. Materials and Methods A UV-visible spectrophotometric technique was developed and validated for the concurrent quantification of TQ and TMZ. This method involved the preparation of standard solutions, determination of maximum absorbance wavelengths, and assessment of linearity, precision, stability, and robustness. Results The validated method exhibited remarkable specificity, linearity, and precision, allowing for accurate absorbance measurements of Thymoquinone (TQ) and Temozolomide (TMZ) without interference from solvents. The method's sensitivity and robustness were validated through low detection limits, high reproducibility, and consistent absorbance patterns across a range of wavelengths and concentrations. Conclusion These findings highlight the method's reliability for the simultaneous quantification of TQ and TMZ.

Keywords: LOD; LOQ; Stability; Temozolomide; Thymoquinone; Validation

FULL TEXT

INTRODUCTION

The most aggressive and dangerous primary tumor of the central nervous system is Glioblastoma (GBM) with a high incidence rate in India. Surgery, radiation, and chemotherapy are the current standard of care for GBM, with TMZ being the standard chemotherapy agent (Araujo et al., 2009). TMZ belonging to class imidazotetrazine, an oral alkylating agent, breaks the dsDNA, leading to cell cycle arrest and ultimately cell death [Chelliah et al., 2021] TMZ [3-methyl-4-oxo-3H,4H-imidazo[4,3-d] tetrazine-8-carboxamide (C6H6N6O2), Mol wt: 194.15] is converted to active form, 5-(3-methyltriazen-1-yl) imidazole-4-carboxamide at physiologic pH (Chowdhury et al., 2018; Fresnais et al., 2021).

However, TMZ faces challenges such as low bioavailability, drug resistance, and systemic toxicity. Additionally, GBM often develops resistance to TMZ, and monotherapy fails to effectively target the heterogeneous tumor cell population. To overcome these challenges, researchers are exploring alternative therapeutic approaches (Gurung, et al., 2010).

Phytochemical compounds, derived from plants, have been extensively explored in recent years for their potential therapeutic applications in treating various diseases like cancer, neuroprotection, anti-inflammatory, antimicrobial, cardiovascular health, etc. (Ismail et al., 2015)

Thymoquinone (TQ), a natural compound derived from the seeds of Nigella sativa, has shown remarkable promise in combating GBM. TQ has been found to intervene with various oncogenic pathways, which are involved in cancer cell growth, proliferation and metastasis. Additionally, TQ has been shown to induce apoptosis, or programmed cell death, in cancer cells, thereby inhibiting tumor growth (Kazi et al., 2018; Khan et al., 2017; Patel et al., 2010; Ramalho et al., 2019; Razak et al., 2013).

The UV visible spectrophotometer is a precise, affordable, and easy method for quantitative analysis of drugs (Shabani et al., 2023) There the no validated methods for the simultaneous analysis of TQ and TMZ by UV spectrophotometry. In current research, we have developed and optimized a UV spectrophotometric method for the quantitative determination of TQ+TMZ which will facilitate future research and quality control.

MATERIALS AND METHODS

Ethical Statement

Not Applicable.

Materials

Thymoquinone and temozolomide were procured from Sigma Aldrich chemicals. Methanol used was analytical grade. A Shimadzu UV- spectrophotometer 1700 (UV-Probe 2.21), with 10 mm quartz cuvettes was used to conduct the analysis (Figure 1).

Figure 1: A. Structure of Temozolomide (Razak et al., 2013). B. Structure of Thymoquinone (Khan et al., 2017).

Preparation of standard solution (10 mg in 10 mL)

The standard solutions were made using equal volume of both the drugs in 10 mL of the solvent (methanol). The solutions were further diluted to serial dilutions Conc range from 2-10 μg/mL.

Optimization and validation

TQ+TMZ, dissolved in methanol showed spectrum with maximum absorbance at 270 nm.

Linearity

The linearity of the method was assessed by measuring the absorbance of TMZ+ TQ at concentrations ranging from 2 to 10 μg/mL, A calibration curve was plotted (concentration vs. absorbance), and statistical parameters like correlation coefficient (R²) and Y-intercept were determined. A similar approach was used by Priya et al., for Apigenin, preparing serial dilutions (2-10 μg/mL) and measuring absorbance at 267 nm in triplicates (Shetti et al., 2022).

Precision

Precision was evaluated through repeatability and intermediate precision. Intraday precision was assessed by measuring absorbance at 270 nm (TMZ + TQ) in five replicates on the same day. Interday precision was determined by analyzing drug concentrations over three different days. % RSD was calculated to assess variability (Singh et al., 2024).

Solution Stability

Solution stability was assessed by preparing fresh stock and dilutions using a fresh solvent. The absorbance of combination (TQ+TMZ) was compared with that of old stock dilutions, and % RSD was calculated to evaluate stability.

Ruggedness and Robustness

Ruggedness was evaluated by preparing five replicates of TQ+TMZ, measuring their absorbance using different instruments, and calculating the % RSD. Robustness was assessed by evaluating the effect of wavelength changes on analytical performance, wherein absorbance was measured at modified wavelengths (TQ+TMZ: 324-270 nm) and the % RSD was calculated (Singhal et al., 2024).

Limit of Detection and Limit of Quantification

Using statistical formulas, LOD and LOQ were calculated.

Statistical Analysis

All measurements were carried out in triplicate and reported as Mean ± Standard Deviation (SD). Linearity was assessed using regression analysis to obtain the calibration parameters, including slope, intercept, and Correlation Coefficient (R²), while method precision, stability, and robustness were evaluated through calculation of percentage relative standard deviation (%RSD). The Limits of Detection (LOD) and Quantification (LOQ) were determined using statistical formulas derived from the calibration curve data.

RESULTS

Optimization and validation

The absorption spectral analysis showed (Figure 2) wavelength at 270 nm.

Figure 2: Spectra of TQ+TMZ.

Linearity

The regression equation yielded a Correlation Coefficient (R2) of 0.995 for TQ+TMZ. By using statistical analysis LOD and LOQ was calculated for TQ+TMZ. The linearity and LOD, LOQ data are summarized in Table 1. While Calibration plot is presented in Figure 3.

Figure 3: Calibration Curve of TQ+TMZ.
Table 1: Linearity of TQ+TMZ.
Sl. No.Conc (µg/mL)Absorbance at 270 nm
120.147
240.33
360.533
480.638
5100.833
R20.99541
LOD0.108434
LOQ0.328588

System Suitability

Analysis of the solvent spectrum revealed no absorbance interference at 270 nm TQ+TMZ there by underscoring the method’s high specificity and selectivity.

Precision

Precision studies revealed that the method is highly reproducible (Interday and Intraday), with %RSD values of less than 2% for 3 replicates measurements of drugs at various precision level, as summarized in Table 2.

Table 2: Precision (Intraday and Interday).
IntradayInterday
ConcMorning 9 amAfternoon 2 pmEvening 5 pmDay 1Day 2Day 3
2 μg/mL0.10.0640.1350.1260.1150.128
2 μg/mL0.10.0650.130.1250.1170.129
2 μg/mL0.090.0660.1380.1270.1180.13
%RSD0.571.5381.2460.7931.3090.775
6 μg/mL0.490.4490.5160.4920.4650.474
6 μg/mL0.4860.4480.5170.4940.4670.473
6 μg/mL0.4890.4490.5180.4960.4690.472
%RSD0.4260.1280.1930.4040.4280.211
10 μg/mL0.7390.6910.740.720.6790.677
10 μg/mL0.7420.6920.7430.7210.680.68
10 μg/mL0.7430.6930.7420.7220.6810.681
%RSD0.280.1440.2050.1380.1470.306

Solution Stability

The stability of TQ+TMZ solution was assessed by comparing the absorbance values of freshly prepared and aged (1 week) solution. The %RSD was calculated, and the results fell within acceptable limits, indicating that the standard stock solution of TQ+TMZ remained stable. The study was conducted for 1 week and data is displayed in Table 3.

Table 3: Solution stability of TQ+TMZ.
Solution StabilityFresh stock dilutionsOld stock dilutions
ReplicatesConc (µg/mL)TQ+TMZTQ+TMZ
160.5220.503
260.5190.509
360.520.51
460.5290.51
560.5250.512
660.5270.513
%RSD0.2930.298

Robustness and Ruggedness

The method was found to be robust, as minor variations in wavelength did not significantly affect the results, and rugged, as %RSD values obtained from replicate analysis using different instruments remained within acceptable limits as shown in Table 4.

Table 4: Robustness and Ruggedness TQ+TMZ.
Robustness TQ+TMZ
Concentration µg/mLAbsorbance at 270 nmAbsorbance at 324 nm
20.120.006
20.1210.007
20.1220.008
%RSD0.82614.285
60.4520.016
60.4540.017
60.4550.018
%RSD0.3365.882
100.6390.008
100.640.009
100.6410.01
%RSD0.15611.111
Ruggedness TQ+TMZ
Concentration µg/mLAnalyst 1Analyst 2
40.330.331
40.3320.332
40.3310.33
%RSD0.30210.3021

DISCUSSION

The UV-spectrophotometric method developed for simultaneous estimation of TQ and TMZ at 270 nm showed good validation results in line with ICH guidelines. The method demonstrated excellent linearity (R²=0.995) in the range of 2-10 µg/mL, with acceptable LOD (0.108) and LOQ (0.328) values indicating adequate sensitivity. No interference was observed at the selected wavelength, confirming specificity. Precision studies, with %RSD values below 2% for both intraday and interday analysis, established the reproducibility of the method. Furthermore, stability testing showed that the TQ+TMZ stock solution remained stable for at least one week. Collectively, these findings confirm that the method is simple, reliable, and suitable for routine analysis of TQ and TMZ in combined formulations.

CONCLUSION

This research showcases the successful development and validation of a novel, UV spectrophotometer for simultaneous estimation of TQ+TMZ detection and quantification in bulk. By harnessing the drug's distinct properties, the proposed method enables the identification of validated parameters, yielding trustworthy and accurate results. Its numerous benefits make it an indispensable tool for quality control, pharmaceutical analysis, and research applications.