Article

Development and Validation of Reverse-Phase High-Performance Liquid Chromatographic Method for Determination of Resveratrol in Human and Rat Plasma for Preclinical and Clinical Studies

Satveer Jagwani1,2, Sunil Jalalpure1,2★, Dinesh Dhamecha2, Gan Siew Hua3, Kiran Jadhav1★ Corresponding author

  1. 1KLE College of Pharmacy, KLE Academy of Higher Education and Research, Nehru Nagar, Belagavi, Karnataka, INDIA.
  2. 2Dr. Prabhakar Kore Basic Science Research Center, KLE Academy of Higher Education and Research, Nehru Nagar, Belagavi, Karnataka, INDIA.
  3. 3School of Pharmacy, Monash University-Malaysia, Jalan Lagoon Selatan, Bandar Sunway, Selangor, MALAYSIA.

CORRESPONDENCE

Sunil Jalalpure

KLE College of Pharmacy, KLE Academy of Higher Education and Research, Nehru Nagar, Belagavi, Karnataka, INDIA.

jalalpuresunil@rediffmail.com

Received: 16-05-2019; Revised: 22-08-2019; Accepted: 02-12-2019.

Volume 54, Issue 1 · pp. 187–193 · PUBLISHED 2020 · DOI: 10.5530/ijper.54.1.22

View on IJPER original site ↗

ABSTRACT

Aim: The aim of the study was to develop and validate a simple and precise reverse-phase High Performance Liquid Chromatography (RP-HPLC) method for quantitative analysis of trans-resveratrol in human and rat plasma. Methods: HPLC method was developed by using Phenomenex Luna C18 column (150 x 4.6 mm, 5 μm) and the optimized mobile phase comprised of acetonitrile/water in isocratic mode (30:70, v/v) with the flow rate of 1.0 mL/min. Trans-resveratrol was detected at a UV wavelength of 306 nm. Developed method was validated as per International Conference on Harmonization (ICH) M10 guidelines. Results: The proposed method was found simple, precise and linear with regression coefficient of 0.999 which could analyse the samples in nanograms levels with mean percent recovery in the acceptable range of 94.44 – 97.44%. The method was precise at the intra-day and inter-day levels as reflected by the relative standard deviation values (less than 3.36%). Trans resveratrol was found to be stable in plasma under different storage conditions. Conclusion: The present investigation demonstrated that the developed method was succesfully applied to accurately determine transresveratrol in human and rat plasma and therefore can be applicable for pre-clinical and clinical studies.

KEYWORDS

Open in new tab

REFERENCES

As published

Showing references and in-text citations exactly as published.

    Cite this article

    SELECT FORMAT

    Jagwani, S., Jalalpure, S., Dhamecha, D., Hua, G. S., & Jadhav, K. (2020). Development and Validation of Reverse-Phase High-Performance Liquid Chromatographic Method for Determination of Resveratrol in Human and Rat Plasma for Preclinical and Clinical Studies. Indian Journal of Pharmaceutical Education and Research, 54(1), 187–193. https://doi.org/10.5530/ijper.54.1.22