Rapid Simultaneous Quantitative Analysis of Hypoglycemic agents by RP HPLC: Development, Validation and Application to Medicine
Mahesh Attimarad1★, Katharigatta Narayanaswamy Venugopala1,2, Mohammed Munirul Islam3, Sheeba Shafi4, Abdulrahman Ibrahim Altaysan1★ Corresponding author
- 1Department of Pharmaceutical Sciences, College of Clinical Pharmacy, King Faisal University, Al Ahsa, SAUDI ARABIA.
- 2Department of Biotechnology and Food Science, Faculty of Applied Sciences, Durban University of Technology, Durban, SOUTH AFRICA.
- 3Department of Biomedical Sciences, College of Clinical Pharmacy, King Faisal University, Al Ahsa, SAUDI ARABIA.
- 4Department of Nursing, College of Applied Medical Sciences, King Faisal University, Al Ahsa, SAUDI ARABIA.
CORRESPONDENCE
Mahesh Attimarad
Department of Pharmaceutical Sciences, College of Clinical Pharmacy, King Faisal University, Al Ahsa, SAUDI ARABIA.
Received: 10-08-2021; Revised: 27-11-2021; Accepted: 11-02-2022.
Volume 56, Issue 2 · pp. 564–572 · PUBLISHED 2022 · DOI: 10.5530/ijper.56.2.79
ABSTRACT
Background: For the treatment of diabetes mellitus type 2, a new formulation containing vildagliptin and remogliflozin was developed. A simple and rapid RP-HPLC method employing linagliptin as an internal standard was developed for quality control of this medicine. Methodology: Formulation analytes, including IS, were separated on a Zorbax C18 column with isocratic elution of acetonitrile and phosphate buffer (pH 5) 55:45 v/v at a flow rate of 1.2 mL/min. The experiment was carried out at room temperature and monitored at a wavelength of 210 nm. The approach was also validated in accordance with the ICH Q2 requirements. Results: The optimized HPLC approach revealed a satisfactory linearity in the concentration ranges of 10-60 μg/mL and 10-100 μg/mL for VIL and REM respectively, with good regression coefficient (R2≥0.998). The average accuracy for VL and REM was 99.57 percent and 100.59 percent, respectively, with a low percentage relative error. The method’s precision was proven by the low percentage relative standard deviation. Furthermore, a robustness assessment employing a Pareto chart generated using a three-level factor interaction study, a multivariate technique, demonstrated that minor changes in individual experimental conditions had no effect on the test results. Finally, the optimized HPLC method was effectively used to assess VIL and Rem from formulation simultaneously. Conclusion: The findings of an assay comparing a simple and rapid isocratic RP-HPLC method devised for the simultaneous quantification of VIL and REM to a previously published approach revealed no significant differences in the assay results. As a result, it might be utilized in any analytical laboratory for quality control of this formulation.
KEYWORDS
REFERENCES
As publishedShowing references and in-text citations exactly as published.
Cite this article
SELECT FORMAT
Attimarad, M., Venugopala, K. N., Islam, M. M., Shafi, S., & Altaysan, A. I. (2022). Rapid Simultaneous Quantitative Analysis of Hypoglycemic agents by RP HPLC: Development, Validation and Application to Medicine. Indian Journal of Pharmaceutical Education and Research, 56(2), 564–572. https://doi.org/10.5530/ijper.56.2.79