Article

A Validated Simple, Rapid and Accurate RP-HPLC Approach for Measuring Vardenafil Hydrochloride Trihydrate in Bulk Drugs and Medicinal Formulations

Ganesh Ramesh Gadekar1★, Rajendra Chandrashekhar Doijad2, Namdeo Ramhari Jadhav2, Somnath Devidas Bhinge3★ Corresponding author

  1. 1Department of Pharmaceutics, Appasaheb Birnale College of Pharmacy-Sangli, South Shivaji Nagar, Sangli, Maharashtra, INDIA.
  2. 2Krishna Institute of Pharmacy, Krishna Vishwa Vidyapeeth (Deemed to be University), Malkapur, Karad, Maharashtra, INDIA.
  3. 3Department of Pharmaceutical Chemistry, Rajarambapu College of Pharmacy, Kasegaon, Walwa, Sangli, Maharashtra, INDIA.

CORRESPONDENCE

Ganesh Ramesh Gadekar

Department of Pharmaceutics, Appasaheb Birnale College of Pharmacy-Sangli, South Shivaji Nagar, Sangli-416416, Maharashtra, INDIA.

gadekarganesh@gmail.com

Received: 09-03-2023; Revised: 04-11-2023; Accepted: 07-03-2024.

Volume 58, Issue 2S · pp. s532–s542 · PUBLISHED Apr-Jun 2024 · DOI: 10.5530/ijper.58.2s.55

View on IJPER original site ↗

ABSTRACT

Background Shortcomings in current analytical methods characterized by time-consuming procedures, limited sensitivity, and high costs associated with reagents and instruments, influence the research background. Objectives The goal of this research was to find a simple, fast, and accurate way to discern and quantify substances in bulk drugs and finished products at the utmost minimal levels of detection and quantification. Materials and Methods The methodology involved using a C18 column then a solvent mixture comprising 0.1% orthophosphoric acid and acetonitrile in an 80:20 proportion delivered at 1 mL/min-1 and observed at 247 nm with a UV detector. Results Following rigorous verification in line with ICH guidelines, the technique revealed several notable features. It revealed rapidity with a retention time of 2.53 min, accurate with average recoveries ranging from 99.20% to 100.43%, precision as demonstrated by a relative standard deviation below 2%, specificity, and a highly linear correlation coefficient of 0.999 between 0.0025 to 6 μg mL-1. Additionally, the method displayed remarkable sensitivity, detecting concentrations as low as 56.91 ng mL-1 and a quantification limit of 172.44 ng mL-1. Conclusion This tested RP-HPLC method has a unique mix of speed, accuracy, and high sensitivity that makes it possible to accurately measure Vardenafil hydrochloride trihydrate, even at nanogram levels. This surpasses established techniques and offers a cost-effective solution for pharmaceutical quality control and analysis.

KEYWORDS

Open in new tab

REFERENCES

As published

Showing references and in-text citations exactly as published.

    Cite this article

    SELECT FORMAT

    Gadekar, G. R., Doijad, R. C., Jadhav, N. R., & Bhinge, S. D. (2024). A Validated Simple, Rapid and Accurate RP-HPLC Approach for Measuring Vardenafil Hydrochloride Trihydrate in Bulk Drugs and Medicinal Formulations. Indian Journal of Pharmaceutical Education and Research, 58(2S), s532–s542. https://doi.org/10.5530/ijper.58.2s.55