Development and Validation of Stability Indicating Liquid Chromatographic Method for Simultaneous Estimation of Silodosin and Dutasteride in API and Combined Dosage Form
Sapna Rathod1★, Ayushi Shah1, Nisarg Patel2, Maulik Mehta3, Bhupendra Prajapati4★ Corresponding author
- 1Department of Pharmaceutical Chemistry and Quality Assurance, APMC College of Pharmaceutical Education and Research, Himatnagar, Gujarat, INDIA.
- 2Department of Pharmacognosy, APMC College of Pharmaceutical Education and Research, Himatnagar, Gujarat, INDIA.
- 3Department of Pharmaceutics, APMC College of Pharmaceutical Education and Research, Himatnagar, Gujarat, INDIA.
- 4Department of Pharmaceutics, Shree S. K. Patel College of Pharmaceutical Education and Research, Ganpat University, Ganpat Vidyanagar, Gujarat, INDIA.
CORRESPONDENCE
Sapna Rathod
Department of Pharmaceutical Chemistry and Quality Assurance, APMC College of Pharmaceutical Education and Research, Himatnagar, Gujarat, INDIA.
Received: 24-02-2025; Revised: 09-05-2025; Accepted: 18-07-2025.
Volume 60, Issue 1S · pp. s312–s319 · PUBLISHED · DOI: 10.5530/ijper.20265613
ABSTRACT
Background:The proposed work was done with a view to develop a RP-HPLC approach for the concurrent quantification of Dutasteride and Silodosin in API and capsule formulation and validate it. Materials and Methods:A C18 Thermo BDS column (150×4.6 mM, 5 μM) was utilized, which also used buffer and acetonitrile as mobile phase (70:30 v/v) eluent A and (30:70 v/v) eluent B at flow rate of 1 mL/min. The samples were quantified at 250 nm employing a PDA detector. In compliance with ICH recommendations, the methodology was validated. Results:For DUT and SIL, linear response was obtained in the range of 5-15 μg/mL and 40-120 μg/mL, individually. The LOD and LOQ values of DUT and SIL were respectively observed 0.302 μg/mL, 1.859 μg/mL and 0.909 μg/mL, 5.634 μg/mL. Both drugs had assay results that ranged from 99.13 to 99.79%. Stress testing using acid, base, peroxide, light and heat were carried out. After adequate treatment, these mixtures were injected in HPLC and it was revealed that products generated from the stress studies were resolved from the peaks of analyte. Conclusion: The current approach was considered accurate, robust, linear, specific and precise and it can be utilized for the concurrent investigation of the drugs stated in its capsule.
KEYWORDS
REFERENCES
As publishedShowing references and in-text citations exactly as published.
Cite this article
SELECT FORMAT
Rathod, S., Shah, A., Patel, N., Mehta, M., & Prajapati, B. (). Development and Validation of Stability Indicating Liquid Chromatographic Method for Simultaneous Estimation of Silodosin and Dutasteride in API and Combined Dosage Form. Indian Journal of Pharmaceutical Education and Research, 60(1S), s312–s319. https://doi.org/10.5530/ijper.20265613